Description
A noninvasive optical method is developed to monitor rapid changes in blood glucose levels in diabetic patients. The system depends on an optical cell built with a LED that emits light of wavelength 535nm that is a peak absorbance of hemoglobin. As the glucose concentration in the blood decreases, its osmolarity also decreases and the RBCs swell and decrease the path length absorption coefficient. Decreasing absorption coefficient increases the transmission of light through the whole blood. The system was tested with a constructed optical cell that held whole blood in a capillary tube. As expected the light transmitted to the photodiode increases with decreasing glucose concentration. The average response time of the system was between 30-40 seconds. The changes in size of the RBC cells in response to glucose concentration changes were confirmed using a cell counter and also visually under microscope. This method does not allow measuring the glucose concentration with an absolute concentration calibration. It is directed towards development of a device to monitor the changes in glucose concentration as an aid to diabetic management. This method might be improvised for precision and resolution and be developed as a ring or an earring that patients can wear.
Download count: 1
Details
Title
- Non-invasive method to detect the changes of glucose concentration in whole blood using photometric technique
Contributors
- Rajan, Shiny Amala Priya (Author)
- Towe, Bruce (Thesis advisor)
- Muthuswamy, Jitendran (Committee member)
- LaBelle, Jeffrey (Committee member)
- Arizona State University (Publisher)
Date Created
The date the item was original created (prior to any relationship with the ASU Digital Repositories.)
2013
Subjects
Resource Type
Collections this item is in
Note
- thesisPartial requirement for: M.S., Arizona State University, 2013
- bibliographyIncludes bibliographical references (p. 46-49)
- Field of study: Bioengineering
Citation and reuse
Statement of Responsibility
by Shiny Amala Priya Rajan