Description
Development of a rapid and label-free Electrochemical Impedance Spectroscopy (EIS) biosensor for Cardiovascular Disease (CVD) detection based on Inerluekin-18 (IL-18) sensitivity was proposed to fill the technology gap between rapid and portable CVD point-of-care diagnosis. IL-18 was chosen for this CVD biosensor due to its ability to detect plaque vulnerability of the heart. Custom (hand) made sensors, which utilized a three electrode configuration with a gold disk working electrode, were created to run EIS using both IL-18 and anti-IL-18 molecules in both purified and blood solutions. The EIS results for IL-18 indicated the optimal detection frequency to be 371Hz. Blood interaction on the working electrode increased the dynamic range of impedance values for the biosensor. Future work includes Developing and testing prototypes of the biosensor along with determining if a Nafion based coating on the working electrode will reduce the dynamic range of impedance values caused by blood interference.
Details
Title
- A rapid and Label-free IL-18 point-of-care biosensor for CVD detection
Contributors
- Jha, Amit (Author)
- LaBelle, Jeffrey (Thesis director)
- Mossman, Kenneth (Committee member)
- Frakes, David (Committee member)
- Barrett, The Honors College (Contributor)
- Harrington Bioengineering Program (Contributor)
- Department of Management (Contributor)
Date Created
The date the item was original created (prior to any relationship with the ASU Digital Repositories.)
2013-05
Resource Type
Collections this item is in